![]() |
|
|
|
|
||
|
Volume 9 Issue 1
|
June 2006
|
![]() |
|
How efficacious are T8 fluorescent lamps?Lamp efficacy is determined by dividing light output by lamp power, resulting in units of lumens per watt (LPW). Determining and comparing lamp efficacy from catalog data is difficult because the actual power needed to operate the lamps is not known (see What are the power requirements of T8 fluorescent lamps?) and because rated light output is given for operation of the lamp at a low frequency of 60 Hz. (T8 lamps are almost always operated on high-frequency electronic ballasts.) A general perspective on the range of T8 lamp efficacy can be gained by dividing the published light output values by some reference power value. For example, NLPIP selected a lamp power of 32.5 watts, which is the standard value listed for this lamp type (ANSI C78.81-2005). Figure 4 shows these efficacy values, which range from 81 to 99 LPW, for 121 T8 fluorescent lamp models from nine manufacturers. It is likely that the upper limit of this range would decrease slightly if it were based on actual power measurements, which were slightly higher than 32.5 watts in this limited study (Figure 3). Nevertheless, the RE70 models tended to be at the low end of the range with all of their efficacies below 88 LPW. Figure 4 also shows lower efficacies for some of the RE80 models at the higher correlated color temperatures (CCT) of 5000 K and 6500 K.
Figure 5 shows the calculated and measured efficacies of the lamps that NLPIP tested. Calculated efficacies were derived by dividing the published light output values by 32.5 watts. NLPIP-measured efficacies were derived from measured light output and power values. The measured efficacies include error bars, which show the combined uncertainty of the measured values for each lamp model. The measured efficacies ranged from 86 to 94 LPW, compared to the calculated efficacies that ranged from 91 to 97 LPW. Both RE80 and RE80 HLO, LL lamps had measured efficacies that were an average of 3.5% lower than the calculated values. This result is due to measured light output values that were lower than the rated values for RE80 lamps and measured lamp power that was higher than 32.5 watts for all lamps.
The efficacies of the RE80 HLO, LL lamps averaged 5.8% higher than the RE80 lamps. As noted in the previous Questions: What is the light output of T8 fluorescent lamps? and What are the power requirements of T8 fluorescent lamps?, both the light output and the input power values of the RE80 HLO, LL lamps were higher than the values for the RE80 lamps. However, the light output values were sufficiently higher to counteract the effect of higher power. The increased efficacies of the RE80 HLO, LL lamps may make them a more attractive choice for new construction or for retrofits in which a ballast change or a luminaire layout change is being made. With a lower ballast factor or a luminaire layout change, customers can take advantage of the higher light output of RE80 HLO, LL lamps; they may be able to meet their illuminance requirements and save energy compared to RE80 lamps. NLPIP testing was conducted using a low-frequency reference ballast as described in American National Standards for Lamp Ballasts-Reference Ballasts for Fluorescent Lamps (ANSI C82.3-2002). T8 lamp efficacy increases by approximately 10% under high-frequency operation (Hitchcock 1983). For more details on testing protocols, see Appendix A: Test methods.
|
|||||
|
|||||
|
||||